Evaluation of the stability of temocillin in elastomeric infusion devices used for outpatient parenteral antimicrobial therapy in accordance with the requirements of the UK NHS Yellow Cover Document.
Fekade Bruck SimeSteven WallisConor JamiesonTim HillsMark GilchristMark SantilloR Andrew SeatonFelicity DrummondJason Robertsnull nullPublished in: European journal of hospital pharmacy : science and practice (2022)
Temocillin reconstituted with 0.3% citrate buffer at pH7 in elastomeric infusion devices can be stored in a fridge (2°C-8°C) for 14 days meeting the YCD acceptance criteria. Considering <5% degradation, the current data supports twice daily dosing of temocillin within the OPAT setting. In jurisdictions where a <10% degradation limit is acceptable, once daily dosing with 24-hour continuous infusion may be considered. Temocillin is a useful alternative to other broad-spectrum anti-Gram-negative agents currently utilised in the OPAT setting and supports the wider antimicrobial stewardship agenda.