Multicolor plate reader fluorescence calibration.
Jacob BealCheryl A TelmerAlejandro VignoniYadira BoadaGeoff S BaldwinLiam HallettTaeyang LeeVinoo SelvarajahSonja BillerbeckBradley BrownGuo-Nan CaiLiang CaiEdward EisensteinDaisuke KigaDavid RossNina AlperovichNoah SprentJaclyn ThompsonEric M YoungDrew EndyTraci Haddock-AngelliPublished in: Synthetic biology (Oxford, England) (2022)
Plate readers are commonly used to measure cell growth and fluorescence, yet the utility and reproducibility of plate reader data is limited by the fact that it is typically reported in arbitrary or relative units. We have previously established a robust serial dilution protocol for calibration of plate reader measurements of absorbance to estimated bacterial cell count and for green fluorescence from proteins expressed in bacterial cells to molecules of equivalent fluorescein. We now extend these protocols to calibration of red fluorescence to the sulforhodamine-101 fluorescent dye and blue fluorescence to Cascade Blue. Evaluating calibration efficacy via an interlaboratory study, we find that these calibrants do indeed provide comparable precision to the prior calibrants and that they enable effective cross-laboratory comparison of measurements of red and blue fluorescence from proteins expressed in bacterial cells.